To ensure patient safety, correct assembling procedure must be followed
Tools required: hexagon screwdriver, torque wrench

1. Hold one side of device securely
2. Secure screws from other side. Repeat for other side (ex: first secure left, then right, or vice versa)

5分11选5NOTE: Torque must be 18kgf-cm ± 1kgf-cm

caution-note

  • Ensure nylock screws are fully tightened

3. DO NOT use the lift scale if screws is loose
4. Screws must be fully locked before use

caution-note-02

caution-note-03

If only one side is fastened, the joint will not be fixed. Usage under these circumstances may result in injury or incidents relating to the lift scale / lift system. Usage is PROHIBITED.